Measurement Time
12.3 Measurement Time
Measurement time differs depending on the measurement conditions. Refer to the following values.
All the values are reference values. Note that they may differ depending on the operating
environment.
12.3.1 LCR Mode
Analog measurement signal (INDEX)
Analog measurement time = A + B + C
Measurement time (EOM)
Measurement time = INDEX + D + E + F + G
A. Analog measurement time
FAST
MED
SLOW
SLOW2
0.5 ms
0.9 ms
2.1 ms
3.7 ms
Tolerance ±0.1 ms
B.
Trigger synchronous output
Trigger synchronous output wait time + INDEX delay time
C. Contact check (DC measurement)
30 µs + 8 µs × Number of WAVEs + DC wait time + AC wait time
Takes double the time if TIMING is BOTH.
D. LCR calculation time
MAX 80 µs
E. Trigger delay time
F.
JUDGE-EOM delay time
G. Judgment
Comparator
MAX 50 µs
BIN
MAX 150 µs
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......