![Hioki IM7580 Скачать руководство пользователя страница 183](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306183.webp)
175
Checking Contact Defects and the Contact State (Contact Check Function)
7.1.4
Monitoring the Detection Level (Detection Level Monitoring
Function)
This function can detect the abnormal measurement waveforms when there is contact between the
test sample and the instrument by monitoring the variations in the r.m.s. value of voltage and the
r.m.s. value of current. During analog measurement, the r.m.s. value of voltage and the r.m.s. value
of current are calculated several times.
The first-calculated r.m.s. value of voltage and the r.m.s value of current are considered as the
reference value respectively. The percentage
∆
% of subsequent r.m.s. values of voltage and
current relative to the reference value is computed using the following formula.
∆
% =
(r.m.s. value - Reference value)
Reference value
× 100 [%]
An error is detected when
∆
% is higher than the set limit value.
1
2
3
1
Press
[SETUP]
.
2
Press
[CONTACT]
tab.
Press
[SWEEP]
tab for the ANALYZER mode.
3
Press
[LEV CHECK]
.
4
5
4
Press
[CHECK]
.
5
Select ON/OFF for the detection level monitoring
function.
[OFF]
Sets the detection level monitoring function to
OFF.
[ON]
Sets the detection level monitoring function to
ON.
7
Application Function
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......