![Hioki IM7580 Скачать руководство пользователя страница 101](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306101.webp)
93
Set Measurement Conditions for Sweep Points
4.4.5
Setting the Delay Time for Each Sweep Point (Point Delay)
Set the delay time for each sweep point in the point delay setting.
In sweep measurement, some measurement samples may require time for the measured value
to stabilize due to transient response. For such cases, set a point delay time.
See “3.2.3 Setting the Delay Time from Trigger to Measurement Start (Trigger Delay)” (p. 34).
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
the last sweep
point
External trigger
Measurement complete
For sequential sweep
Trigger delay
Point delay
For repeat sweep
Measurement of
sweep point 1
Measurement of
sweep point 2
Returns to the measurement of sweep point 1.
Internal Trigger
Measurement of
sweep point 3
Measurement of
the last sweep
point
For step sweep
Measurement complete
Measurement complete
Measurement complete
Measurement complete
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
the last sweep
point
External trigger
External trigger
External trigger
2
1
The numeric keypad can be
used for input.
1
Use
/
to enter the delay time.
(With the numeric keypad, press
[SET]
.)
Settable range 0.00000 s to 9.99999 s
[C]
Sets to the default value. (0.00000 s)
2
Press
[EXIT]
to close the setting screen.
3
Press
[EXIT]
to close the advanced settings
screen.
4
Analyzer Function
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......