4.3 Setting DC Resistance Measurement
82
In cases such as the following, set the time to wait before starting DC resistance measurement.
• When measurement is changed from AC signal measurement to DC resistance measurement.
• When the voltage level is changed.
• When DC adjustment is enabled.
This delay time is for delaying measurement until the DC level and 0 VDC has stabilized.
When measuring DC resistance, observe the measurement waveform and then insert the delay time required
until the transient phenomenon disappears because the time required until the DC level stabilizes differs
depending on the test sample.
4.3.7 Setting the Delay Time
When number of averaging times is 1
Setting changed
•
Changed from AC measure-
ment to DC measurement
•
Signal level changed
•
Range changed
DC delay
(Initial value: 0.0003
seconds)
Measure DC
resistance
DC delay
(Initial value: 0.0003
seconds)
Measure offset
When DC adjustment is enabled (ON)
End measurement
When the number of averaging times is 2 or more
(The number of times is 2 in this example)
Setting changed
•
Changed from AC mea-
surement to DC mea-
surement
•
Signal level changed
•
Range changed
DC delay
(Initial value:
0.0003 seconds)
Measure DC
resistance
DC delay
(Initial value:
0.0003 seconds)
Measure offset
When DC adjustment is enabled (ON)
DC delay
(Initial value:
0.0003 seconds)
Measure DC
resistance
Measure DC
resistance
DC delay
(Initial value:
0.0003 seconds)
Measure
offset
Measure
offset
When DC adjustment is enabled (ON)
Obtain
average
value
Measurement for set number
of averaging times
End mea-
surement
Measurement for set number
of averaging times
Procedure
LCR Initial Screen
1
Rdc Basic Settings
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......