Appendix 8 Selecting the Equivalent Circuit Model
A12
When using the equivalent circuit function, it is important to select an appropriate equivalent circuit model.
The following table provides examples of measurement targets and equivalent circuit models:
Since the models for which parameters can be accurately acquired varies with observed values, perform a
simulation for estimated results and select the equivalent circuit model based on a comparison with observed
values.
When automatically selecting the equivalent circuit model, it will not be possible to select the optimal model
when the acquisition of frequency characteristics fails to yield local extreme values. Consequently, set the
sweep range so that resonance characteristics can be accurately acquired.
Appendix 8
Selecting the Equivalent Circuit Model
Measurement target
Corresponding equivalent
circuit model
Inductor
Inductor with high core loss and low ESR
A
Comparatively high ESR
B
Capacitor
Significant leak resistance effect
C
Typical capacitor
D
Resistor
Low resistance value, significant inductance effect
B
High resistance value, significant stray capacitance effect
C
Piezoelectric element
-
E
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......