13.3 About Measurement Times and Measurement Speed
399
13
Chapte
r 1
3 S
pec
ifications
Measurement times differ depending on the measurement conditions. Refer to the following values.
When there is an external trigger, the corresponding time shown in the table below is added to the analog
measurement signal.
* Low Z high accuracy mode is only valid when the frequency is between 4 Hz and 100 kHz.
13.3 About Measurement Times and
Measurement Speed
All of the values are reference values. Note that they may differ depending on the conditions of use.
Analog measurement signal (INDEX)
FAST
MED
SLOW
SLOW2
DC
1 ms+0.6 ms
10 ms+0.6 ms
100 ms+0.6 ms
400 ms+0.6 ms
4.00 Hz to 19.99 Hz
Tf+0.6 ms
2
×
Tf+0.6 ms
3
×
Tf+0.6 ms
4
×
Tf+0.6 ms
20.00 Hz to 100.00 Hz
Tf+0.6 ms
2
×
Tf+0.6 ms
8
×
Tf+0.6 ms
64
×
Tf+0.6 ms
100.01 Hz to 999.99 Hz
Tf+0.6 ms
4
×
Tf+0.6 ms
16
×
Tf+0.6 ms
128
×
Tf+0.6 ms
1.0000 kHz to 3.0000 kHz
Tf+0.2 ms
8
×
Tf+0.2 ms
32
×
Tf+0.2 ms
256
×
Tf+0.2 ms
3.0001 kHz to 10.000 kHz
Tf+0.2 ms
16
×
Tf+0.2 ms
64
×
Tf+0.2 ms
512
×
Tf+0.2 ms
10.001 kHz to 30.000 kHz
4
×
Tf+0.2 ms
64
×
Tf+0.2 ms
256
×
Tf+0.2 ms
2048
×
Tf+0.2 ms
30.001 kHz to 100.00 kHz
8
×
Tf+0.1 ms
128
×
Tf+0.1 ms
512
×
Tf+0.1 ms
4096
×
Tf+0.1 ms
100.01 kHz to 300.00 kHz
32
×
Tf+0.1 ms
512
×
Tf+0.1 ms
2048
×
Tf+0.1 ms
16384
×
Tf+0.1 ms
300.01 kHz to 1.0000 MHz
64
×
Tf+0.1 ms
1024
×
Tf+0.1 ms
4096
×
Tf+0.1 ms
32768
×
Tf+0.1 ms
1.0001 MHz to 5.0000 MHz
400
×
Tf+0.1 ms
6400
×
Tf+0.1 ms
25600
×
Tf+0.1 ms
204800
×
Tf+0.1 ms
Tolerance: ±5% ±0.2 ms Tf [s]=1 ÷ measurement frequency
The DC time is the time when DC adjustment is OFF. The time is about twice as long when DC adjustment is ON.
Range
Level
Frequency
Low Z High Accuracy Mode
DC Bias
Time
100 m
, 1
All levels
DC, 4 Hz to 999.99 Hz
ON
OFF
5 ms
100 m
, 1
All levels
1 kHz to 5 MHz *
ON
OFF
1 ms
100 m
, 1
All levels
ALL frequency *
ON
ON
1 ms
100 m
, 1
All levels
DC, ALL frequency
OFF
OFF
1 ms
10
0.005 V to 0.1 V
DC, ALL frequency
OFF
OFF
1 ms
10
0.101 V to 1 V
DC, 4 Hz to 1 MHz
OFF
OFF
1 ms
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......