13.1 General Specifications
387
13
Chapte
r 1
3 S
pec
ifications
Specifications
Chapter 13
13.1 General Specifications
1.Basic Specifications
Measurement mode
(1) LCR mode: Measurement with single condition
(2) Analyzer mode: Measurement frequency sweep, measurement level sweep
• Measurement points: 2 to 801
• Sweep method: Normal sweep/segment sweep
Normal sweep: START-STOP/ CENTER-SPAN/ START-STEP, maximum 801 points
Segment sweep: START-STOP, maximum 20 segments (total of 801 points)
• Display: List display/graph display
(3) Continuous measurement mode: Consecutive measurements with saved conditions
• LCR mode: Up to 30 conditions
• Analyzer mode: Up to 2 conditions
* Continuous measurements including a mixture of LCR mode and analyzer mode are
possible.
Measurement items
Z (Impedance), Y (Admittance),
(Phase angle), Rs (Equivalent series resistance (ESR)),
Rp (Parallel equivalent resistance), X (Reactance), G (Conductance), B (Susceptance),
Ls (Series equivalent inductance), Lp (Parallel equivalent inductance), Cs (Series equivalent
capacitance), Cp (Parallel equivalent capacitance), Q (Q factor), D (Loss coefficient tan
,
Rdc (DC resistance)
Display range
Z, Y, Rs, Rp, Rdc, X, G, B, Ls, Lp, Cs, Cp:
±
(0.000000 (unit) to 9.999999G (unit))
* Absolute value display for Z and Y only
:
±
(0.000
°
to 999.999
°
)
D:
±
(0.000000 to 9.999999)
Q:
±
(0.00 to 99999.99)
%:
±
(0.0000% to 999.9999%)
* DISP OUT is displayed when the upper limit is exceeded.
Measurement frequency
(1)Frequency range
4 Hz to 5 MHz
(2)Setting resolution
5-digit resolution (however, the minimum resolution is 10 mHz)
(3)Frequency accuracy
±
0.01%
Output impedance
(Hc terminal, when 1 kHz)
Normal mode: 100
±
10
Low impedance accuracy mode: Approx. 10
(100 m
and 1
ranges of 100 kHz or less)
Approx. 100
(Except for the above)
* The following impedances are connected to the Lc terminal.
1
range or less (100 kHz or less) when low impedance high accuracy mode: 5
or less
1 V or less of other than the above:
3 k
range or more...............Approx. 35
1 k
range or less...............Approx. 5
1.001 V or more of other than the above:
30 k
range or more...............Approx. 35
10 k
range or less...............Approx. 5
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......