![Hioki IM3570 Скачать руководство пользователя страница 310](http://html1.mh-extra.com/html/hioki/im3570/im3570_instruction-manual_2137305310.webp)
7.5 Calculating Values Scaling
302
Scaling applies a compensation function to the measurement value.This function can be used to provide com-
patibility among measurement devices. Set the compensation coefficients a and b for the measurement val-
ues of the first to fourth parameters to compensate by the following expression.
See
"Appendix 1 Measurement Parameters and Arithmetic Expressions"(p. A1)
However, if the parameter corresponding to X is either D or Q, scaling is applied to
as shown in the following
expression, and then D or Q is obtained from
'.
7.5 Calculating Values Scaling
LCR
ANALYZER
Y = a × X + b
' = a ×
+ b
X: the first to fourth parameter measurement value
a: integration value of the measured value X
Y: the last measurement value
b: the value added to measured value X
': compensation value of
Procedure
Compensation Screen
1
LCR Initial Screen
2
Select ,
and
press
to
confirm the selection.
When you want to cancel scaling:
1.
Press
to switch to the [Compensation Screen].
2.
Press
and select
.
3.
Press
to confirm the setting.
Scaling Compensation Setting
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......