6.5 Setting Application Settings
270
This section describes how to set the trigger. In continuous measurement mode, measurement will be per-
formed in accordance with the trigger settings configured as described in this section. Two types of triggers
can be set: sequential measurement and step measurement.
6.5.2 Setting the Trigger
Procedure
Continuous Measurement Initial Screen
Application Settings
1
Press
.
2
Application Settings
Select the trigger setting.
Performs sequential measurement.
When the external trigger is input, all panels
set to
on the setting screen are
measured.
Performs step measurement.
When the external trigger is input, the panel
set to
on the setting screen is mea-
sured, and the instrument switches to the
next panel’s measurement conditions.
3
Trigger Setting
4
Press
to close the setting screen.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......