5.8 Editing Judgment Points
218
Measurement parameter
Settable range
Frequency
4 Hz to 5 MHz
Open circuit voltage level
0.005 V to 5.000 V
Voltage level between test
sample terminals
0.005 V to 5.000 V
Current level between test
sample terminals
0.01 mA to 50 mA
3
Set each of the following settings.
Edit Sweep Point
Sweep Point & Upper/Lower Limit Value Settings
6
Press
to close the setting screen.
Press
.
Set the sweep point.
Set the upper limit value of the comparator.
Set the lower limit value of the comparator.
4
Sweep Point Setting
Sweep Point Setting
Use the numeric keypad to enter the setting of the sweep point and press a unit key to confirm the
setting.
5
Press
to close the setting screen.
When you want to cancel the setting:
Press
.
Settable range
Settable range
• Sweep point:
Measurement parameter
Settable range
Frequency
4 Hz to 5 MHz
Open circuit voltage level
0.005 V to 5.000 V
Voltage level between test
sample terminals
0.005 V to 5.000 V
Current level between test
sample terminals
0.01 mA to 50 mA
• Upper limit value: -
9.999999G to 9.999999G
• Lower limit value: -
9.999999G to 9.999999G
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......