![Hioki IM3570 Скачать руководство пользователя страница 209](http://html1.mh-extra.com/html/hioki/im3570/im3570_instruction-manual_2137305209.webp)
5.7 Judging Measurement Results (Comparator Function)
201
5
Chapte
r 5
ANAL
YZER Functi
o
n
When the
setting is
Upper/Lower Limit Value Settings
Set the judgment area.
Sets the upper and lower limit values
based on the current measurement
values.
Sets the reference value, upper limit
value, and lower limit value.
Sets the upper and lower limit values
as percentage values relative to the
reference value.
Sets the upper and lower limit values
as absolute values relative to the refer-
ence value.
When you want to cancel the setting:
Press
.
Set the upper and lower limit values based on the current
measurement values.
If you make a mistake during input:
press
to cancel the input and start again.
1.
Press
and use the numeric keypad to set the
upper limit value.
• Settable range: -999.9999 to 999.9999
(Set as a % value)
• Settable range: -9.999999G to 9.999999G
(Set as a reference value)
2.
Press
to confirm the upper limit value.
3.
Press
and use the numeric keypad to set the
lower limit value.
• Settable range: -999.9999 to 999.9999
(Set as a % value)
• Settable range: -9.999999G to 9.999999G
(Set as a reference value)
4.
Press
to confirm the lower limit value.
5.
Press
to confirm the setting.
Upper/Lower Limit Value Settings
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......