5.3 Normal Sweep
145
5
Chapte
r 5
ANAL
YZER Functi
o
n
Setting the start value and end value of the sweep
1.
Press
and use the numeric keypad to set the
start value of the sweep.
Settable range: 4 Hz to 5 MHz
2.
Press
,
, or
key to confirm the setting.
3.
Press
and use the numeric keypad to set the
end value of the sweep.
Settable range: 4 Hz to 5 MHz
4.
Press
,
, or
key to confirm the setting.
5.
Press
and use the numeric keypad to enter the
number of sweep points.
Settable range: 2 to 801
6.
Press
key to confirm the setting.
7.
Select the setting method for sweep points (when fre-
quency sweep).
START-STOP Settings
The sweep points are calculated linearly from
the setting values of
,
,
and .
The sweep points are calculated logarithmical-
ly from the setting values of
,
, and
.
8.
Press
to confirm the setting.
If you make a mistake during input:
press
to cancel the input and start again.
• The setting method of the sweep points can only be selected when the sweep parameter is fre-
quency and the setting method of the sweep range is
. In other cases, the setting
method of the sweep points is fixed to linear.
• If the setting method of the sweep points is changed, the horizontal axis scale of the graph display
screen changes as shown in the figures below. (The horizontal axis scale of the graph can also be
changed with the horizontal axis scale setting.)
See:
"Horizontal Axis Scale Setting" (p. 177)
Linear Setting
Log Setting
The horizontal scale is linear display.
The horizontal scale is log display.
When the
setting is
.
When the
setting is
.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......