77
Judging Measurement Results
Absolute value setting
Set the value after setting the judgment mode (p. 70) to
BIN
.
1
Touch the
BIN
key on the measurement
screen.
2
Touch the
Z
key.
The key display differs depending
on the measurement parameter
3
Touch the
ABS
key and then touch the
EXIT
key.
2
1
4
Use the
key to select the BIN num-
ber to set, and touch the
EDIT
key.
1
2
5
Touch the
HI
key for the No. 1 param-
eter.
No. 1 parameter
No. 3 parameter
6
Enter the upper limit value with the nu-
meric keypad.
Settable range: -9.99999 G to 9.99999 G
When you do not want to set the upper and
lower limit values, touch the
OFF
key.
7
Touch the
ENTER
key to confirm the up
-
per limit value.
The display will return to the screen shown
in step 5.
8
Touch the
LO
key for the No. 1 parameter,
set the lower limit value with the numeric
keypad, and then touch the
ENTER
key.
Settable range: -9.99999 G to 9.99999 G
The display will return to the screen shown
in step 5.
9
Using the same procedure, set the up-
per and lower limit values for the No. 3
parameter.
10
Touch the
EXIT
key twice.
Displays the measurement screen.
3
Performing Measurements in LCR Mode
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...