72
Judging Measurement Results
Absolute value setting
Set the value after setting the judgment mode (p. 70) to
COMP
.
This explanation uses the example of setting the measurement conditions for the No. 1 parameter.
1
Touch the
LMT
key on the measurement
screen.
2
Touch the
ABS
key.
3
Touch the
HI
key, and set the upper limit
value with the numeric keypad.
2
1
×10
3
Step the units up.
×1/10
3
Step the units down.
Units: a/ f/ p/ n/ µ/ m/ none/ k/ M/ G
Settable range: -9.99999 G to 9.99999 G
If you do not wish to set an upper limit
value, touch the
OFF
key.
4
Touch the
ENTER
key to confirm the up
-
per limit value.
The display will return to the screen shown
in step 2.
5
Touch the
LO
key, set the lower limit
value with the numeric keypad, and
touch the
ENTER
key.
Settable range: -9.99999 G to 9.99999 G
If you do not wish to set an lower limit
value, touch the
OFF
key.
6
Touch the
EXIT
key.
Displays the measurement screen.