54
Setting Measurement Conditions (basic settings)
Line frequency
(DC)
When performing DC measurement, be sure to set the line frequency of the power supply being
used.
Screen display method (For more information, see p. 24.):
(Measurement screen)
SET
key>(
SET
screen)
Rdc
tab>
LINE FREQ
key
1
Select the line frequency.
50 Hz
Sets the line frequency to 50 Hz.
60 Hz
Sets the line frequency to 60 Hz.
2
Touch the
EXIT
key twice.
Displays the measurement screen.
To suppress noise, the instrument needs to be set to match the frequency of the power source. Before
operating, set the instrument to the frequency of your commercial power. If the supply frequency is not set
properly, measurements will be unstable.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...