52
Setting Measurement Conditions (basic settings)
Constant current (CC) mode setting
Low Z high accuracy mode
Low Z high accuracy mode
Constant current setting range
0.01 mA to 50.00 mA
0.01 mA to 100.00 mA
Constant current accuracy
±1%±10 µA
Output impedance
100
Ω
±10
Ω
10
Ω
±2
Ω
For some samples, you may not be able to perform constant-current measurement (measurement in con
-
stant-current mode). In this case, the following error will be displayed:
Constant current measurement will not be performed. Change the constant-current value to a value that is
less than or equal to the value displayed for Iac.
(Example: Constant-current measurable range when measuring an L value of 1 mH at 1 kHz)
Sample impedance
Zm
becomes as follows:
Zm = Rm + jXm = 0 [
Ω
] -
j6.28 [
Ω
]
Xm = 2
π
fL
The impedance
Zm’
as seen from the instrument’s voltage generator is as follows:
Zm’ = Ro + Zm = 100 [
Ω
] -
j6.28 [
Ω
] Ro:
Output resistance (100 [
Ω
])
Accordingly, the current
Im
across both leads of the sample is as follows:
Im =
Vo
|
Zm’
|
=
Vo
100.2 [
Ω
]
V
o
:
generator output
Since the instrument’s voltage generator output range is 10 [mV] to 5 [V] as per the above table, the con
-
stant-voltage measurable range is
Im
= 0.10 [mA] to 49.9 [mA] based on the above formula.
When low Z high accuracy mode is enabled, the output resistance
Ro
will be 10 [Ω].
•
When the measurement value is outside the guaranteed accuracy range, the following error massage ap
-
pears at the top of the screen.
In this case, you should consider the following possible causes, and you should either change the mea
-
surement signal level and measurement range while checking the accuracy assured ranges "Guaranteed
accuracy measurement level range" (p. 213), or you should consider the measured values as values for
reference.
•
Measurement signal level is too low: Increase the test signal level.
•
The current measurement range is inappropriate (when using the HOLD setting): set again in the AUTO
•
range, or change the range by manual.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...