233
Error Massage and Error Display
Error
overview
Error
Description
Solution and reference
Contact check error
The letter A (“after”) will be shown
if the contact check error occurred
after measurement, while the let
-
ter B (“before”) will be shown if
the contact check error occurred
before measurement.
The unconnected terminals will be
indicated.
HL : H
POT
, H
CUR
, L
POT
, L
CUR
L : L
POT
, L
CUR
H : H
POT
, H
CUR
•This message will be dis-
played in the measured
value display area if one of
the terminals is found to be
disconnected, for example
due to a broken wire, after
measurement.
Check the connection of each
terminal.
•The sample has not been
secured properly in the mea-
surement cable, probe, or
fixture.
Check the manner in which
the sample has been secured.
Check the wiring for wiring
breaks or poor contact.
See the probe or fixture’s user
check function (detecting poor
contact with the sample during
•This message will also be
displayed when using the
instrument in an environment
characterized by a large
amount of noise.
Consider taking the following
measures:
•Use guarding.
•
Implement anti-external
noise measures.
of External Noise" (p. Appx.5).
•
Separate the sample, mea
-
surement cables, and instru
-
ment from the source of the
noise (motor, inverter, electro
-
magnetic switch, power line,
equipment generating sparks,
etc.) or perform the measure
-
ment in a separate room.
•Plug the instrument into a
grounded outlet.
•Use a separate power sup-
ply from the device that is
generating the noise.
Fan motor error
•
The fan motor has stopped
or is operating at low speed.
•
Turn off the instrument and
verify that its air vents are
unobstructed and free of any
foreign material.
•
If no issues are found, the
instrument needs to be
repaired. Contact your au-
thorized Hioki distributor or
reseller.
Maintenance and Service
11
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...