178
Internal Circuitry
Electrical specifications
Input signals
Input type
Optocoupler-isolated, non-voltage contact inputs
(current sink, active-low)
Input asserted (on) voltage
0.9 V or less
Input de-asserted (off) voltage
Open or 5 V to 24 V
Input asserted (on) current
3 mA/ch
Maximum applied voltage
30 V
Output signals
Output type
Isolated by photo-couplers, Nch open drain output
(current sink, active-low)
Maximum load voltage
30 V
Maximum output current
50 mA/ch
Residual voltage
1 V or less (50 mA)
Internally isolated
power supply
Output voltage
4.5 V to 5.0 V
Maximum output current
100 mA
External power input
none
Connection examples
Input circuit connection examples
IM3536
IM3536
IM3536
IM3536
Input
Switch connections
Input
Relay connections
Input
PLC output (sink output) connections
Output
Common
Common
Input
Output
PLC output (source output) connections