4.3 Setting DC Resistance Measurement
87
4
Chapte
r 4
LCR Function
When performing DC resistance measurement, be sure to set the line frequency of the power supply being used.
4.3.4
Setting the Line Frequency
It is necessary to set the line frequency for the power supply being used so that the instru-
ment can reject noise. Set this parameter to the frequency of the commercial power supply
being used before using the instrument to make measurements. Failure to properly set the
line frequency will prevent you from acquiring stable measurement values.
Procedure
LCR Measurement Screen
1
Rdc Basic Settings
Press .
2
Rdc Basic Settings
Select the line frequency.
Sets the line frequency to 50 Hz.
Sets the line frequency to 60 Hz.
3
Line Frequency Setting
4
Press
to close the setting screen.