4.2 Setting Basic Settings of Measurement Conditions
74
With the averaging function, the measured values can be averaged. Using this function, it is possible to
reduce fluctuations in the measured value display.
When the number of averaging times is 4, the number of measurements, measurement output points, and
measurement value calculation method during output are as follows.
4.2.8 Displaying Average Values (Averaging Set)
With internal trigger
A rolling average of the tested values over the set number of times for
averaging is always calculated backwards from the present.
(
When the sample to be tested is changed over, it takes a little time for a certain stabili-
zation time period until the results is reliable.)
With external trigger
Average over the number of averaging times based on trigger input.
M1
M2
M3
M4
M5
M6
M7
M8
M9
M10
M1
1
M1+M2
2
M1+M2+M3
3
M1+M2+M3+M4
4
M2+M3+M4+M5
4
M3+M4+M5+M6
4
M4+M5+M6+M7
4
M5+M6+M7+M8
4
M6+M7+M8+M9
4
M7+M8+M9+M10
4
Measurement points
Measurement value output
Moving average
Time
M1
M2
M3
M4
M5
M6
M7
M8
M9
M10
Time
Arithmetic mean
M1+M2+M3+M4
4
M5+M6+M7+M8
4
Measurement points
Measurement value output
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......