8.5 Calculating Values (Scaling)
246
Scaling applies a compensation function to the measurement value.This function can be used to provide com-
patibility among measurement devices. Set the compensation coefficients a and b for the measurement val-
ues of the first to fourth parameters to compensate by the following expression.
In transformer mode, the compensation coefficients a and b are set relative to the calculation parameter cal-
culated value, and compensation is applied using the following expression:
See
"Appendix1 Measurement Parameters and Calculation Formula"(p. A1)
However, if the parameter corresponding to X is either D or Q, scaling is applied to
as shown in the following
expression, and then D or Q is obtained from
'.
8.5 Calculating Values (Scaling)
Y = a × X + b
' = a ×
+ b
X: the first or third parameter measurement value
a: integration value of the measured value X
Y: the last measurement value
b: the value added to measured value X
': compensation value of
Procedure
Compensation Screen
1
LCR measurement Screen
2
Select ,
and
press
to
confirm the selection.
When you want to cancel scaling:
1.
Press
to switch to the [Compensation Screen].
2.
Press
and select
.
3.
Press
to confirm the setting.
Scaling Compensation Setting
Содержание IM3533
Страница 2: ......
Страница 8: ...Contents vi...
Страница 52: ...3 3 When TRANSFORMER Mode 44...
Страница 156: ...4 5 Setting Application Settings 148...
Страница 200: ...5 3 Application Settings 192...
Страница 216: ...6 4 Setting Application Settings 208...
Страница 270: ...9 4 Deleting a Panel 262...
Страница 316: ...11 7 File and Folder Operations 308...
Страница 338: ...12 6 Measurement Using a Computer 330...
Страница 396: ...Appendix 13 Device Compliance Statement A26...
Страница 399: ......
Страница 400: ......