2.6 Advanced Settings and Functions
54
This section describes how to initialize the instrument’s settings.
All measurement data (up to 2,000 values) will be deleted.
To disable APS temporarily
Turn on the instrument by pressing the power key while
holding down the
HOLD
key to disable APS until the
next time the instrument’s power is cycled. The next
time the power is cycled, APS will be enabled (as long
as the APS setting is enabled in sub-function mode).
Reverting the instrument to factory settings
(system reset)
1.
Press the power key while holding down the
Fn
key.
The instrument will enter sub-function mode.
2.
Using the
▼
and
▲
keys, select the System Reset screen
and press the
OK
key.
The
[
OK?
]
mark will flash.
3.
Press the
OK
key again.
The instrument will revert to the factory settings.
• If the System Reset screen is displayed by mistake,
cycle the instrument's power without pressing the
OK
key. Instrument operation will be restored without a
system reset having been performed.
• For more information about how to clear previously saved
measurement data, see “Clearing stored data (p.44)”
.
HIOKI FT6380B981-01