67
4
Maintenance and Service
4.1 Repair, Inspection, and Cleaning
DANGER
Customers are not allowed to modify, disassemble, or
repair the instrument.
Doing so may cause fire, electric shock, or injury.
Calibrations
IMPORTANT
Periodic calibration is necessary in order to ensure that the
instrument provides correct measurement results of the specified
accuracy.
The calibration frequency varies depending on the status of the
instrument or installation environment. We recommend that the
calibration frequency is determined in accordance with the status of
the instrument or installation environment and that you request that
calibration be performed periodically.
Cleaning
•
To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent.
When rinsing with water, make sure that the battery cover has
been attached. Make sure that the protector has been removed
and rinse with room-temperature water.
•
After cleaning, thoroughly wipe off water of the device and
protector. Completely dry them before putting the protector on.
•
Rinse with room-temperature water. If rinsing with hot water, the
waterproof performance may be impaired.
•
Do not put water while the battery cover has been taken off. It
may cause failure due to water intrusion.