87
Measurement Principle
6.2
Measurement Principle
While applying a voltage of AC power supply between H(C)
electrode and E electrode, the flowing AC current
I
is measured with
an ammeter. In addition, the voltage
V
between S(P) electrode and
E electrode that is caused by the flow of the current
I
is measured
by an AC voltmeter.
The earth resistance
R
X
of E electrode is derived from the measured
current
I
and voltage
V
. It is not possible to accurately measure the
voltages between H(C) and E electrodes and between H(C) and
S(P) electrodes.
Voltmeter
AC Power
supply
l
I
V
=
R
X
×
I
R
X
=
V
/
I
E electrode
S(P) electrode
H(C) electrode
Ammeter