83
Error Display
5.3 Error Display
Error
display
Description
Solution
Err 0
Ω
ADJ
Zero adjustment is outside
the allowable range, zero
adjustment has been executed
while it is not on HOLD.
Execute zero adjustment
again in accordance
with the zero adjustment
Err1
Abnormality in firmware
Instrument failure.
Contact your authorized
Hioki distributor or reseller
to organize repair.
Err2
Adjustment data damaged
Err4
Setting data damaged
Err5
Abnormality in measurement
circuit
Err8
Z3210 communications error
(connection failure; Z3210 or
hardware failure)
Take the following actions:
•
Reinstall the Z3210.
•
Install a different Z3210.
If the error persists, you are
experiencing a instrument
failure. Contact your
authorized Hioki distributor
or reseller to organize
repair.
APS
→P.oFF
Instrument powered off by
APS
Cycle the power.
bAtt
→P.oFF
Instrument powered off due to
supply voltage drop
HIOKI FT6031B961-00