81
Accuracy table
(4)
AC+DC voltage
Measured value/MAX/MIN/AVE (CM4371
-50
)
Range
(automatic
ranging
threshold)
Accuracy
guarantee range
(resolution)
Accuracy guarantee
frequency range
Measurement accuracy
Filter disabled
Filter enabled
20.00 A
(more
than 2000
counts)
1.00 A to 20.00 A
(0.01 A)
10 Hz ≤ f < 45 Hz
±1.8% rdg ±0.10 A
±2.3% rdg ±0.10 A
DC, 45 Hz ≤ f ≤ 66 Hz
±1.3% rdg ±0.13 A
±1.8% rdg ±0.13 A
66 Hz < f < 1 kHz
±2.0% rdg ±0.10 A
−
600.0 A
(less than
180 counts)
1.0 A to 600.0 A
(0.1 A)
10 Hz ≤ f < 45 Hz
±1.8% rdg ±0.7 A
±2.3% rdg ±0.7 A
DC, 45 Hz ≤ f ≤ 66 Hz
±1.3% rdg ±1.3 A
±1.8% rdg ±1.3 A
66 Hz < f < 1 kHz
±2.0% rdg ±0.7 A
−