71
Input Specifications, Measurement Specifications
Effects of
conductor
position*
1
Cable diameter
Measurement
area
Accuracy
Measurement
area diagram
8 mm
2
cross-linked
polyethylene insulated vinyl
sheath cable
(Finished outer diameter:
8.6 mm)
Area A
Within ±3.0% rdg
B
A
Area B
Within ±7.0% rdg
8 mm
2
cross-linked
polyethylene insulated vinyl
sheath cable
(Finished outer diameter: 13
mm)
Area A
Within ±2.0% rdg
Area B
Within ±5.0% rdg
Effects of radiated
radio-frequency
electromagnetic
field
Add ±2% rdg at 10 V/m.
*1: Defined assuming that 100 A current having a frequency of 55 Hz is measured, with respect to the jaw’s
center-point.