86
External Input and Output Terminals and Signals
Signal functions
Isolated power supply
Pin number
Signal name
EXT. I/O MODE switch (NPN/PNP) setting
NPN
PNP
8
ISO_5V
Isolated power 5 V
Insulated power supply −5 V
9, 27
ISO_COM
Isolated power supply common
Input signal
Pin number
Signal name
Function
1
START
Starts a test, and generates output voltage.
20
STOP
Ends a test, and shuts down output voltage.
3
INTERLOCK
Cancels the interlock state.
The interlock state is canceled while the signal is at the same potential
(ON) as ISO_COM.
Placing the signal in the open (OFF) state triggers the interlock,
preventing testing from starting.
The test is stopped when it turns OFF during the test.
See “2.6 Shutting Down the Output of the Instrument (Interlock
4, 5, 22, 23
LOAD0 to LOAD3
Selects the number of the panel to be loaded.
6
LOAD_VALID
Executes the panel load.
[INFO:002 Panel loading...]
is displayed during panel load and does
not accept any key input.
Output signal
Pin number
Signal name
Function
30
PASS
Outputs when the comparator judgment is PASS.
11
UPPER FAIL
Outputs when the comparator judgment is UPPER FAIL.
12
LOWER FAIL
Outputs when the comparator judgment is LOWER FAIL.
29
C_CHECK_FAIL
Outputs when the contact check judgment is FAIL.
13
BDD
Outputs when the BDD count is 1 or more.
28
TEST
Outputs during a test.
31
VON
Outputs when the voltage monitor value is within
±
10% of the set
voltage value.
10
SYSTEM_ERR
Outputs when the following errors occur.
•
When an internal temperature error occurs
•
When a fan error occurs
•
When a model information error occurs
•
When a setting backup error occurs
EXT. I/O input and output signals cannot be used while the measurement conditions are being
changed inside the instrument.
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...
Страница 215: ...www hioki com HIOKI BT5525A961 00 ...