66
Limiting the Current to Be Applied to the Device Under Test (DUT)
Setting the current limit value automatically
1
Open the Setting screen.
[SET]
Select
2
Set
[CURRENT LIMIT]
to
[AUTO]
.
Select
[AUTO]
Select
3
Set charging time
[CHARGE TIME]
.
Select
[EDIT]
Edit
•
Change the value (up/down keys)
•
Change the cursor position
(left/right keys)
Confirm
4
Set
[DUT CAPACITY]
.
Select
[MANUAL]
[AUTO]
[EXIT]
Return to the Measurement screen
IMPORTANT: Charging time
Charging may take longer than the set charging time (
[CHARGE TIME]
) depending on the DUT’s
capacitance component. This can be addressed by either of the following steps:
•
Set a shorter charging time (
[CHARGE TIME]
).
•
Set the
[CURRENT LIMIT]
to
[MANUAL]
and adjust the current limit value to achieve the
desired charging time.
Factors which can cause charging to take longer than the set charging time include the
instrument’s internal load and the current limit circuit’s response time.
HIOKI BT5525A961-00