144
Communications Command Reference
Setting the capacitance value of the DUT
Syntax
Command
:CHARge:CAPacity <Capacitance value of the DUT>
•
Capacitance value of the DUT {0.1E-09 to 200.0E-09}
Sets the capacitance value of the DUT.
Description
Specifies the capacitance value of the DUT.
The capacitance value of the DUT is used to calculate the current limit value.
When this command is executed, the auto capacitance value setting is OFF (MANUAL).
Example
:CHARge:CAPacity 99E-09
The capacitance value of the DUT is set to 99 nF.
Querying the capacitance value of the DUT
Syntax
Query
:CHARge:CAPacity?
Response
(9 bytes +
CRLF)
<Capacitance value of the DUT>
•
Capacitance value of the DUT {0.1E-09 to 200.0E-09}
Returns the capacitance value of the DUT.
Description
Queries the capacitance value of the DUT.
Example
:CHARge:CAPacity?
␣
99.0E-09
The capacitance value of the DUT is 99 nF.
Setting the auto capacitance value of the DUT
Syntax
Command
:CHARge:CAPacity:AUTO {ON|OFF}
Description
Allows the capacitance value of the DUT to be set automatically.
ON
The capacitance value of the DUT is measured and the capacitance value is
automatically set at the start of the test.
OFF
The capacitance value of the DUT is not automatically set and is fixed.
See “5.3 Limiting the Current to Be Applied to the Device Under Test (DUT)” (p. 64).
Example
:CHARge:CAPacity:AUTO ON
The auto capacitance value of the DUT is set to ON (AUTO).
Querying the auto capacitance value of the DUT
Syntax
Query
:CHARge:CAPacity:AUTO?
Response
(2 or 3 bytes +
CRLF)
{ON|OFF}
Description
Queries the auto capacitance value of the DUT.
Example
:CHARge:CAPacity:AUTO?
ON
The auto capacitance value of the DUT is ON (AUTO).
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...
Страница 215: ...www hioki com HIOKI BT5525A961 00 ...