139
Communications Command Reference
Querying the voltage monitor value
Syntax
Query
:MEASure:MONItor?
Response
(3 bytes +
CRLF)
<Voltage monitor value>
•
Voltage monitor value {25 to 500}
Returns the voltage monitor value.
Description
Queries the voltage monitor value.
Example
:MEASure:MONItor?
␣
35
The voltage monitor value is 35 V.
Clearing the measured value
Syntax
Command
:MEASure:CLEar
Description
Clears the measured value and judgment result. The judgment is determined to be no
judgment.
Example
:MEASure:CLEar
Clears the measured value and judgment result.
Querying the number of measured values saved in the memory
Syntax
Query
:MEASure:COUNt?
Response
(3 bytes +
CRLF)
<Number of measured values in the memory>
•
Number of measured values saved in the memory {0 to 999}
Returns the number of measured values saved in the memory.
Description
Queries the number of measured values saved in the memory.
As the test starts, the memory is cleared once and then the measured values are saved in
the internal memory for each sampling. Up to 999 measured values can be saved in the
memory and any measured values from samplings taking place after that are not saved.
By querying the number of measured values saved in the memory, you can find out the
number of measured values returned with the
:MEASure:MEMory?
query.
Example
:MEASure:COUNt?
␣
23
23 measured values are saved in the memory.
Querying the measured values saved in the memory
Syntax
Query
:MEASure:MEMory?
[CRLF]
Response
<Measured value 1>,<Measured value 2>,<Measured
value 3>,
<Measured value N>
Description
Queries all the measured values saved in the memory.
The measured values are returned with comma delimiters. When
CRLF
is specified as an
option, they are returned with line break (
CRLF
) delimiters.
The number of measured values (N), can be checked with the
:MEASure:COUNt?
query.
The response description is specified by
:MEASure:VALid
See the
:MEASure?
query for the measured value response format.
An execution error occurs if the measured values are not saved in the memory.
Example
:MEASure:MEMory?
120.0E+06,123.4E+06,124.5E+06,130.9E+06,150.0E+06
Five measured values are saved in the memory and the respective resistance values are
120.0 M
Ω
, 123.4 M
Ω
, 124.5 M
Ω
, 130.9 M
Ω
, and 150.0 M
Ω
.
Communications Function
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...
Страница 215: ...www hioki com HIOKI BT5525A961 00 ...