111
Automatically Sending the Measured Value Every Time a Test Is Completed (Auto Data Output Function)
8.5 Automatically Sending the Measured
Value Every Time a Test Is Completed
(Auto Data Output Function)
After the test is complete, the measured value can be automatically sent to an external device via
the instrument's communications interface.
IMPORTANT
When using the auto data output function, do not send a command to the instrument until the
measured value is received after the test starts. The measured value may be sent in duplicate.
1
Open the Setting screen.
[SET]
Select
2
Set the interface to which data is automatically output by
[AUTO DATA OUT]
.
Select
[OFF]
(Not sending automatically)
[RS232C]
[USB]
[LAN]
3
Select the data type.
Select
[TYPE1]
Sends automatically in
the format of type 1
[TYPE2]
Sends automatically in
the format of type 2
[EXIT]
Return to the Measurement screen
See “Communications settings” (p. 167) “
:SYSTem:COMMunicate:DATAout
command”.