3.7 Displaying Measurement Results
40
• A high contact resistance or a high wiring resistance may cause an increase in
measurement errors.
Hioki cannot guarantee the measurement accuracy if the sum of contact resis
-
tance and wiring resistance reaches 40
(4
for the 3 m
and 30 m
ranges).
• The instrument may not detect measurement malfunctions if the measure
-
ment lead has a capacitance of 1 nF or more.
Overflow is indicated by “
OF
” or “
-OF
” on the display, caused by one of the fol
-
lowing:
BT3561A, BT3562A, BT3563A
Range
SOURCE H-L
SENSE H-L
3 m
5.5
6.5
30 m
5.5
6.5
300 m
15
30
3
150
30
30
1.5 k
150
300
6 k
2.5 k
3000
6 k
25 k
Overflow Display
Display
Condition
OF
• The measured value exceeds the limit of the current
measurement range
• The test object impedance exceeds the input level.
• When the result of relative value calculation is larger
than +99.999%.
-OF
• The measured value is below the limit of the current
measurement range
• The test object impedance exceeds the input level (in
the negative direction).
• When the result of relative value calculation is smaller
than -99.999%.
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...