8.6 Message Reference
139
8
C
ha
pte
r 8
R
S-2
32
C
/G
P-IB
/L
AN
In
te
rfa
ce
s
Set and Query the Comparator Decision Tolerance Setting
(Comparator Function)
(Resistance Measurement)
Syntax
Command
:CALCulate:LIMit:RESistance:PERCent
<Tolerance (%)>
Query
:CALCulate:LIMit:RESistance:PERCent?
Response
<Tolerance (%)>
<Tolerance (%)>
= 0 to 99.999
(NR2)
Example
Command
:CALC:LIM:RES:PERC 0.3
Query
:CALC:LIM:RES:PERC?
Response
0.300
(Voltage Measurement)
Syntax
Command
:CALCulate:LIMit:VOLTage:PERCent
<Tolerance (%)>
Query
:CALCulate:LIMit:VOLTage:PERCent?
Response
<Tolerance (%)>
<Tolerance (%)>
= 0 to 99.999
(NR2)
Example
Command
:CALC:LIM:VOLT:PERC 1.538
Query
:CALC:LIM:VOLT:PERC?
Response
1.538
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...