7.1 Making Input Waveform Display Settings (Analog Waveforms)
173
7
Chap
ter 7 W
ave
form
Display Se
ttings
Description
Magnification and compression (p. 212) in the voltage axis direction is based on
the zero position.
Although the range of voltage that can be displayed on the Waveform screen
depends on the zero position and magnification/compression of the voltage axis,
the measurement range is unaffected.
0V
Zero Position: 0%
Zero Position: 50% Zero Position: 100%
(Magnification: x 1)
Measurement Range
Display Range
0%
100%
50%
0%
100%
0%
100%
50%
50%