6.6 Setting Trigger Timing
144
Set waveform recording operation when a trigger event occurs.
Timing for the Recorder function is set as follows.
Timing for the Memory function can be selected by various trigger settings.
Description
About trigger timing
The selected trigger mode determines how recording stops.
6.6
Setting Trigger Timing
Trigger Timing Setting
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Trigger Settings screen
See
Screen Layout (p. 35)
Operating Key
Procedure
1
CURSOR
Move the cursor to the
[Trigger Timing]
item.
2
F1 to F8
Select either choice.
Start
Start recording when a trigger event occurs,
and stop after the specified recording
length.(default setting)
Stop
Start recording when the START key is
pressed, and stop when a trigger event oc-
curs.
Start &
Stop
Record the interval from one trigger event un-
til the next trigger event.
(Select either Start or Stop triggering for each
channel on the Analog and Logic pages.)
[Start]
[Stop]
[Start & Stop]
T
Recording
START key
T
Recording
T
T
Recording
Recording
Starts
Recording starts when a trig-
ger event occurs
Recording starts when you
press the START key
Recording starts when a
Start trigger event occurs
Recording Stops
With
[Single]
trigger mode
Recording stops after data
has been acquired for the
specified recording length
Recording stops when a trig-
ger event occurs
Recording stops when a Stop
trigger event occurs
If no trigger event occurs before the specified recording length elapses:
[Start]
,
[Stop]
or
[Start & Stop]
: Recording stops after data has been acquired for the specified
recording length
With
[Repeat]
trigger mode
The Trigger Wait state begins
after data has been acquired
for the specified recording
length
When another trigger event
occurs, data is again acquired
for the specified recording
length, then Trigger Wait re-
sumes (repeats)
When a trigger event occurs,
recording stops and then
starts again (repeats)
When a trigger event occurs,
recording stops and the Trig-
ger Wait state resumes
When another trigger event
occurs, recording continues
until the next trigger occurs
(repeats)
If no trigger event occurs before the specified recording length elapses:
[Stop]
: After data is acquired for the specified recording length, recording restarts. This repeats
until a trigger event occurs.
[Start & Stop]
: The Trigger Wait state begins after data has been acquired for the specified re-
cording length (Start Trigger)
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......