5.4 Converting Input Values (Scaling Function)
126
Scaling Setting Examples
Using a Clamp-On Probe
To set automatically
To set manually (enter two-point numerical values)
The 9018-50 Clamp-On Probe provides 0.2 V output when measuring 10 A. So
Scaling should be set to display 10 A with 0.2 V input (and 0 A with 0 V input).
However, you may need to switch the vertical axis (voltage range) to suit actual
input values.
For example, to display ±0.2 V at full scale, set the vertical display to 20 mV per
division (the instrument’s 20 mV/div range)
Example 1.
Measure with the 10A range of the Model 9018-50 Clamp-On Probe and
display the measured data in units of [A] (Amperes)
Select the clamp type to be
used.
1
Select the
F1
[Clamp]
button.
2
(When the range is displayed)
Select the same range as that set
with the clamp.
The Clamp Scaling dialog appears.
Select the
[Exec]
button.
Scaling is performed automatically to
suit the selected clamp.
3
4
Setting Items
Setting Choice
Scaling
On
Format
Decimal
Units
A
(Scaling Method)
2-Point
Input 1
Scale 1
0.2
10
(0.2:displays as 200m)
Input 2
Scale 2
0
0
Inputs 1 and 2:
[V] value
Physical values 1 and 2: [A] value (value
of displayed measurement units)
With scaling, signals from the sensor are
acquired as current values.
A/B cursors and gauges are displayed
and printed with current (Ampere) val-
ues.
See
"Applying Gauges" (p. 198)
"List and Gauge Settings" (p. 330)
Before Scaling
After Scaling
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......