4.2 Setting Measurement Configuration (Status Settings Screen)
90
About timebase and sampling setting
The timebase setting establishes the rate of input signal waveform acquisition,
specified as time-per-division on the horizontal axis (time/div).
The sampling setting specifies the interval from one sample to the next. For
details about sampling, refer to "Appendix 4.1 Sampling" (p. A48).
4.2.2 Setting the Timebase (Horizontal Axis) and Sam-
pling Rate
Function
Explanation
Memory
Function
• The timebase and sampling rate settings are interdependent.
Changing the timebase causes the sampling rate to be changed. The number of samples
per division is fixed at 100. Therefore, the sampling period is 1/100
th
of the timebase set-
ting.
• If the appropriate timebase setting for the input signal is unknown:
Set the timebase automatically.
"3.3.5 Automatic Range Setting (Auto-Ranging Function)" (p. 74)
• To acquire waveforms with different sampling rates for each channel:
Set different sampling rates for Timebase 1 and Timebase 2. Set Timebase 2 to the slower
sampling rate.
"Setting Different Sampling Rates" (p. 94)
• Using the Model 8958 16-Ch Scanner Unit:
If other input modules are installed together with the Model 8958, the other modules are
set to Timebase 1, and the 8958 to Timebase 2.
"Setting Timebase 1 and 2: When using the Model 8958 16-Ch Scanner Unit together with
If only the Model 8958 is installed, it is set to Timebase 1.
"Setting Channels to Use: When using only the Model 8958 16-Ch Scanner Unit" (p. 89)
• Setting the sampling period according to an external signal: (External Sampling)
"14.2.3 External Sampling (EXT.SMPL)" (p. 382)
Recorder
Function
The timebase and sampling rate can be set independently.
The sampling rate (from 100 ns/S to 1 s/S) is selected depending on the timebase setting.
See
"Appendix 4.4 Recorder Function Values" (p. A50)
REC&MEM
Function
• Memory Waveform: The timebase and sampling rate settings are linked.
• Recorder Waveform: The timebase and sampling rate settings can be set independently.
However, this sampling rate and that of the Memory Waveform are one and the same.
Real-Time
Saving
Function
• The timebase and sampling rate settings are interdependent. Changing the timebase
causes the sampling rate to be changed. The number of samples per division is fixed at
100. Therefore, the sampling period is 1/100
th
of the timebase setting.
• The timebase for the whole waveform can be set automatically. This selects the most
suitable timebase according to the measurement waveform timebase set for real-time
data and the selected save destination. When set manually, the timebase can be selected
from 10 ms/div to 1 hour/div.
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......