17.3 Trigger Section
300
17.3 Trigger Section
Trigger method
Digital comparison
Trigger modes
•
Memory Function, FFT Function: Single, Repeat, or Automatic
•
Recorder Function: Single or Repeat
Trigger source
Analog Unit (Ch1 to Ch16), Standard Logic 16 Ch Logic Unit (Max. 3 units, 48 Channels)
External Trigger occurs by applying a 2.5 V falling edge signal, or shorted terminals
Independent trigger criteria settable for each channel
Free-run operation occurs when all trigger types are off.
Manual trigger
Included
Trigger criteria
AND or OR of each trigger source
Trigger types (analog)
• Level Trigger
Set digitally as a voltage value below full-scale
• Voltage Sag Trigger (Drop)
Triggering occurs when peak voltage falls below the specified level (for commercial power).
(for 50/60 Hz commercial power)
• Window Trigger
Upper and lower trigger threshold levels are specified
Triggering occurs when the signal enters or exits the defined threshold range.
• Period Trigger
A trigger period reference voltage level and period range are specified
The period of the signal rising (or falling) through the specified level is measured, and triggering
occurs when the period is outside of the specified range.
• Glitch Trigger
Set voltage level and pulse width (glitch width)
Triggering occurs when the signal pulse width is narrower than the specified pulse width defined
as rising or falling through a specified voltage level.
• Event Trigger
Triggering occurs when the event values specified for the level or glitch trigger are exceeded.
Triggering occurs when the input rises above (or falls below) the specified trigger voltage level.
Trigger types (logic)
Pattern (mask) trigger by 1, 0, or x (x: don't care)
Trigger filter
•
Memory Function, FFT Function : Off, 0.1, 0.2, 0.5, 1.0, 1.5, 2.0, 2.5, 5.0, 10.0 div
•
Recorder Function
: Off/On (10 ms fixed)
Trigger types (external trigger)
External signal is used as trigger.
Trigger types (timer trigger)
Triggering occurs at a preset time.
Trigger level resolution
0.1% f.s. (f.s. = 20 div)
Pre-trigger
Memory Function, FFT Function
% setting: 0, 2, 5, 10, 20, 30, 40, 50, 60, 70, 80, 90, 95, 100, -95%
div setting: Setting by units of 1 div
Trigger priority
Off/On
Trigger timing
Start, Stop and Start & Stop (Recorder Function)
Trigger output
Open-collector output (with 5 V output, Active Low)
Pulse width at level setting time: (sampling rate × no. of data points after trigger) or more
Pulse width at pulse setting time :2 ms ± 1 ms
Trigger input and output termi-
nals
Terminal Block
Level display function
Included
Trigger marker
Trigger marks indicate positions of trigger events
Trigger search function
Searches for positions that meet trigger criteria
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......