13.3 Setting FFT Analysis Conditions
224
The display range of the vertical (y) axis can be set to automatically suit analysis results, and can be freely
expanded and compressed.
13.3.10Setting the Display Range of the Vertical Axis
(Scaling)
1
2
1
Select automatic or manual scaling of the y-axis display.
Move the cursor to the
[Scale]
column of the Analysis No. to set.
Select
2
When
[Manu]
is selected
Set the upper and lower limits to display.
Move the cursor to the
[Lower]
or
[Upper]
item.
Set the upper and lower limits to display the analysis results.
Setting range: -9.9999E+29 to +9.9999E+29
(with exponent from E-29 to E+29)
See:
Auto
Scaling of the vertical (y) axis is automatically set according to
analysis results. (default setting)
Manu-
al
Scaling of the vertical (y) axis can be set as desired, to suit the pur-
pose of the measurement.
This is useful for magnifying or reducing the displayed amplitude,
and for shifting the displayed waveform up or down.
Procedure
To open the screen: Press the
STATUS
key
→
[Status]
sheet
How do I copy settings to other calculation No.?
See:
"8.9 Copying settings to other channels (calculation No.) (Copy function)" (
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......