9.2 Setting the Trigger Mode
153
9
Chapt
er
9
T
rigger
Set
tings
Set whether to continue to accept triggers after measuring.
If all trigger sources are disabled (Off, with no trigger setting), measurement starts immediately
(free-running).
Description
Available selections depend on the function.
To Stop Measuring:
Press the
STOP
key.
Press once: recording stops at the end of the specified recording length.
Press twice: recording stops immediately.
When the trigger mode is set to
[Repeat]
During the processing interval from the end of recording until entering the next
trigger standby condition (Auto Save, Auto Print, waveform display processing,
numerical calculation), triggering will not be accepted.
9.2
Setting the Trigger Mode
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
1
Move the cursor to the
[Trigger]
item.
2
Select the trigger mode.
Select
Single
Only one trigger is recognized. After pressing the
START
key, once a trigger is applied, a waveform is recorded for
the specified recording length, and measurement then
stops.
Repeat
Triggers are accepted continuously.
When no trigger is applied, the instrument enters the Trig-
ger Wait state. Press the
STOP
key to stop measuring.
(See below)
Auto
Triggers are accepted continuously.
If no trigger is applied within about one second, a waveform
of the specified recording length is automatically recorded.
Press the
STOP
key to stop measuring.
Trigger Mode
Function
Single
O
O
(default setting)
Repeat
O
O
Auto
O
(default setting)
×
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......