8.10 Making Detailed Settings for Input Modules
149
8
Chapt
er 8
Ut
ilit
y Funct
ions
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (
Mode
Switches between voltage measurement and RMS measurement.
Select
Response
Response can be set to three speeds: Fast, Normal and Slow.
Normally set to
[Fast]
, this can be changed to
[Normal]
or
[Slow]
to stabilize the
display when measuring low frequencies, or when severe fluctuations are
present.
Select
8.10.7 Settings for the 8972 DC/RMS Unit
Selections
Description
DC
Voltage measurement (default setting)
RMS
RMS measurement
Selections
Description
Fast
Sets the response time to about 100 ms. (default setting)
Normal
Sets the response time to about 800 ms.
Slow
Sets the response time to about 5 s.
Содержание 8847
Страница 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Страница 2: ......
Страница 18: ...Operating Precautions 8 ...
Страница 26: ...1 4 Basic Key Operations 16 ...
Страница 74: ...4 4 Waveform Observation 64 ...
Страница 98: ...5 5 Managing Files 88 ...
Страница 110: ...6 6 Miscellaneous Printing Functions 100 ...
Страница 126: ...7 8 Seeing Block Waveforms 116 ...
Страница 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Страница 198: ...10 6 Numerical Calculation Type and Description 188 ...
Страница 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Страница 304: ...16 2 External I O 294 ...
Страница 362: ...Appendix 5 FFT Definitions A28 ...
Страница 368: ...Index Index 6 ...
Страница 369: ......
Страница 370: ......
Страница 371: ......