2.1 Pre-Operation Inspection
18
7.
(Confirming blown fuse for mA, mA terminal)
Set the function switch to "Diode Check (
)".
Connect the red test lead to the µA, mA terminal,
and the black test lead to the V terminal. Short the
tips of the red and black test leads by touching
them together.
8.
(Confirming blown fuse for A terminal)
Connect the red test lead to the A terminal, and
the black test lead to the V terminal. Short the tips
of the red and black test leads by touching them
together.
This procedure only partially confirms the opera-
tion of this instrument. Periodic calibration is nec-
essary in order to ensure that this instrument
operates according to its product specifications.
•
Approximately 1.6 V displayed
→
OK
•
OL
displayed
→
NO
Fuse may be blown
.
Please replace fuse.
•
Buzzer sounds
→
OK
•
OL
displayed
→
NO
Fuse may be blown.
Please replace fuse.
Содержание 3804-50
Страница 2: ......
Страница 12: ...ご使用にあたっての注意 8 ...
Страница 18: ...1 2 各部の名称と機能 14 スタンドを立てたまま 上方向から強い力を加えないでくださ い スタンドを損傷します ネジ 通信ポート 背面 ...
Страница 30: ...2 7 コンデンサ容量測定 26 ...
Страница 42: ...3 11 通信機能 オプション 38 コネクタを本器から外す場合はツメを押しながら引き 抜きます ツメ ...
Страница 54: ...5 3 確度 50 ...
Страница 60: ...6 5 全点灯表示の確認方法 56 ...
Страница 64: ......
Страница 65: ...TEMPERATURE 3415 02 XXXX 製品名 HiTESTER 3415 01 3415 INSTRUCTION MANUAL 3804 50 DIGITAL HiTESTER ...
Страница 66: ......
Страница 76: ...8 ...
Страница 94: ...2 7 Capacitance Measurement 26 ...
Страница 106: ...3 11 Communications Function 38 ...
Страница 125: ...索引 4 3 2 1 7 6 5 10 9 8 ...
Страница 126: ......
Страница 129: ......