8.9 Message Reference
202
Setting and Query of Short Circuit Compensation Function
Syntax
Command
Query
:CORRection:SHORt
<Character>
<Character>
= ALL/ ON/ OFF/ RETurn
:CORRection:SHORt?
<Character>
= ALL/ ON/ SPOT/ OFF
Explanation
Command Sets the short circuit compensation function.
If ALL or ON is set, short circuit compensation value begins to be
incorporated and then short circuit compensation is enabled after
incorporation ends.
ALL
:
Delete disabled short circuit compensation value, acquire
short circuit compensation value according to the measure-
ment values set at short circuit compensation reading set-
ting command (:CORRection:SHORt:POINt) and enable the
short circuit compensation function.
ON
:
Delete disabled short circuit compensation value, acquire
short circuit compensation value at the current measure-
ment conditions (frequency, level) and enable the short cir-
cuit compensation function.
OFF
:
Disable short circuit compensation value at the enabled
measurement conditions (frequency, level).
RETurn
:
Enable all disabled short circuit compensation values.
Query
Returns the setting of the short circuit compensation function as char-
acters.
ALL : The short compensation function is enabled for all measure-
ment conditions (frequency, level).
ON
: The short compensation function is enabled for the current
measurement conditions (frequency, level).
SPOT : The short compensation function is enabled for other than the
current measurement conditions (frequency, level).
OFF : The short compensation function is disabled.
Example
Command
:CORRection:SHORt ON
Delete disabled short circuit compensation value, acquire short circuit
compensation value at the current measurement conditions (fre-
quency, level) and enable the short circuit compensation function.
Query
Response
:CORRection:SHORt?
:CORRECTION:SHORT ON
(when HEADER ON)
ON
(when HEADER OFF)
The short compensation function is enabled for the current measure-
ment conditions (frequency, level).
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......