4.5 Current Integration
71
1
2
3
4
5
6
7
8
9
10
M
easurement
P
ro
cedure
Output Settings
If no integral value output is needed, skip ahead to "Starting Mea-
surement" (page 73).
11.
Confirm that the optional 9094 OUTPUT CORD is inserted
securely into output jack 1 (OUT1)
(OUT2 is for integral value out-
put)
.
12.
Press
SHIFT
→
(
OUTPUT
) to display the output settings.
13.
In DC mode, select the integral value for output by pressing
SHIFT
→
(
OUTPUT
).
On the display,
OUT2
is fixed to
REC
(12-bit D/A output of integral value: 1 mV
per count up to ± 999.9 mV).
Output setting is disabled when neither OUT1 or OUT2 is displayed.
Mode Setting
Bar Graph Display
OUT2: REC
DC
±
Total integral output
+
Positive integral output
-
Negative integral output
AC, AC+DC
none
Integral output (fixed)
• No polarity sign (±, + or -) is displayed in AC and AC+DC
modes.
• Select the appropriate current range before making output set-
tings.
❖
See "Integral Range Setting" (page 70)
• Each five counts increments or decrements the output by 0.5
mV (0.5 mV resolution).
• During repeating measurement, the integral value output is
reset to 0 V at the end of each measurement interval.
• See "Output Settings" (page 54) for settings applicable to the
OUT1 jack.
• The refresh rate of the integral value (D/A) output is the same
as the display update rate.
This rate is 100 ms when FAST is selected.
Содержание 3290-10
Страница 1: ...CLAMP ON AC DC HiTESTER INSTRUCTION MANUAL 3290 10...
Страница 2: ......
Страница 6: ...iv Contents...
Страница 14: ...8 Usage Notes...
Страница 42: ...36 2 5 Turning Power On and Off...
Страница 54: ...48 3 6 Measuring Duty by Timer...
Страница 149: ...4 5 6 7 8 9 10 11...
Страница 150: ......