6.3 CLAMP ON AC/DC SENSOR Specifications
6.3.1 Model 9691 CLAMP ON AC/DC SENSOR
6.3 CLAMP ON AC/DC SENSOR
Specifications
Rated current
AC/DC 100 A
Output voltage
AC/DC 1 V /100 A
Output resistance
100
Ω
or less
Maximum
permissible input
100 Arms continuous (Derating according to frequency)
Maximum peak
current value
150 A
Continuous input
❖
See "Continuous Input" (page 117)
Peak input
❖
Frequency band
DC to 10 kHz (-3dB)
Effect of conductor
position
Within ± 1.0% (at 80 A (55 Hz))
Effect of external
electromagnetic field
0.5 A equivalent or less
(in an external electromagnetic field of 400 A/m)
Temperature
characteristics
0 to 40
°
C range: 0.1 x accuracy specifications/
°
C
32 to 104°F range: 0.18 x accuracy specifications/°F
Location for use
Indoor, altitude up to 2000 m (6566 feet)
Operating
temperature and
humidity range
0 to 40°C (32 to 104°F), 80% RH or less (no condensation)
Storage temperature
and humidity
-10 to 50°C (14 to 122°F), 80% RH or less (no condensation)
Operating temperature
and humidity for
guaranteed accuracy
Guaranteed
accuracy period
23 ±
5°C (73 ± 9°F), 80% RH or less (no condensation)
For one year (Opening and closing of the sensor: 10,000 times)
Maximum permitted
circuit voltage
600 V
Dielectric strength
5.55 kV AC for 1 minute (sensor - case, sensor - circuit)
Power supply
±
5 V ±
0.25 V (supply from the outside)
Maximum rated power
50 mVA
Measurable conductor
diameter
35 mm (1.38") or less
Содержание 3290-10
Страница 1: ...CLAMP ON AC DC HiTESTER INSTRUCTION MANUAL 3290 10...
Страница 2: ......
Страница 6: ...iv Contents...
Страница 14: ...8 Usage Notes...
Страница 42: ...36 2 5 Turning Power On and Off...
Страница 54: ...48 3 6 Measuring Duty by Timer...
Страница 149: ...4 5 6 7 8 9 10 11...
Страница 150: ......