Warranty

Warranty malfunctions occurring under conditions of normal use in conformity
with the Instruction Manual and Product Precautionary Markings will be repaired
free of charge. This warranty is valid for a period of three (3) years from the date
of purchase.  Please contact the distributor from which you purchased the prod-
uct for further information on warranty provisions.

Introduction

Thank you for purchasing the HIOKI Model 3288-20 CLAMP ON AC/DC HiT-
ESTER. To obtain maximum performance from the instrument, please read this
manual first, and keep it handy for future reference.

Initial Inspection

When you receive the instrument, inspect it carefully to ensure that no damage
occurred during shipping. If damage is evident, or if it fails to operate according
to the specifications, contact your dealer or Hioki representative.

The HIOKI Model 3288-20 CLAMP ON AC/DC HiTESTER is a compact and
lightweight instrument that enables you to measure up to maximum of AC/DC
1000 A. Besides measuring current, the 3288-20 also contains DMM functions
for AC and DC voltage, resistance and continuity testing. With "True RMS" mea-
surement, it can handle measurement of distorted current waveforms. In AC cur-
rent mode it guarantees accuracy of measurement for frequency from 10 Hz.

I

This manual contains information and warnings essential for safe operation of
the instrument and for maintaining it in safe operating condition. Before using it,
be sure to carefully read the following safety precautions. 

Safety Symbols

Symbols for Various Standards

The following symbols in this manual indicate the relative importance of cautions
and warnings.

Measurement categories 

This instrument the current measurement section complies with CAT III 600 V safe-
ty requirements, and the voltage measurement section complies with CAT III 300
V, CAT II 600 V safety requirements. To ensure safe operation of measurement in-
struments, IEC 61010 establishes safety standards for various electrical environ-
ments, categorized as CAT II to CAT IV, and called measurement categories.

CAT II: Primary electrical circuits in equip-
ment connected to an AC electrical outlet
by a power cord (portable tools, house-
hold appliances, etc.) CAT II covers direct-
ly measuring electrical outlet receptacles.
CAT III: Primary electrical circuits of heavy
equipment (fixed installations) connected
directly to the distribution panel, and feed-
ers from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution panel).
Using a measurement instrument in an environment designated with a higher-
numbered category than that for which the instrument is rated could result in a
severe accident, and must be carefully avoided.
Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV mea-
surement applications could result in a severe accident, and must be carefully
avoided.

Follow these precautions to ensure safe operation and to obtain the full benefits
of the various functions.
Before using the instrument the first time, verify that it operates normally to en-
sure that no damage occurred during storage or shipping. If you find any dam-
age, contact your dealer or Hioki representative.

The display is blanked automatically. (Auto Power Save Function)

• The auto power save function is activated automatically when the power is

turned on. (Not possible to cancel)

• This function automatically switches to the power save state when 10 min-

utes have elapsed since the last operation. (Power save state)

• To restore from the auto power save state, turn the function selector to the

OFF position once.

Zero-adjust Function

• Before measuring DC current [

A], you must perform zero adjustment by

simultaneously pressing the 

 and 

HOLD

 keys while there is no input to

the instrument.

• The zero adjustment function compensates for sensor magnetization and

changes in current display over time. This function is only effective with mea-
surement of DC current [

A].

The measurement range is automatically set to the most appro-
priate range. (Auto-range Function)

When measuring an AC current [

A], DC current [

A], AC voltage [

V],

DC voltage [

V], or resistance [

], the measurement range is automatically

set to the most appropriate range. 

A manual range setting becomes available. (Manual Range Function)

Power on the tester while holding down the 

 or 

HOLD

 key to select a

manual range for measuring AC current [

A], DC current [

A], AC voltage

[

V], DC voltage [

V] or resistance [

]. Note that this function is not avail-

able for continuity testing. Press the 

key to step to the next range. 

To switch between AC voltage [

V] and DC voltage [

V], press and hold

the 

 key for at least one second. 

Indication when the input is out of range. (Overflow indication)

When the input exceeds the measurement range, "O.F" or "-O.F" is displayed. 

Accuracy

• Guaranteed accuracy period is 1 year (Opening and closing of the Clamp sen-

sor 10,000 times, whichever comes first).

• Accuracy  guarantee for temperature and humidity: 23±5°C (73°F±9°F) and

80%RH or less (no condensation)

• Battery warning indicator is not lighting. 

AC current measurement: true RMS value, DC current measurement: average value

Overview

Safety

This instrument is designed to comply with IEC 61010 Safety
Standards, and has been thoroughly tested for safety prior to
shipment. However, mishandling during use could result in
injury or death, as well as damage to the instrument. Using the
instrument in a way not described in this manual may negate
the provided safety features. Be certain that you understand the
instructions and precautions in the manual before use. We dis-
claim any responsibility for accidents or injuries not resulting
directly from instrument defects.

3288-20

CLAMP ON AC/DC HiTESTER

Instruction Manual

February 2015  Revised edition 6

  Printed in Japan

3288C981-06  15-02H

In the manual, the 

 symbol indicates particularly important information

that the user should read before using the instrument.
The 

 symbol printed on the instrument indicates that the user should

refer to a corresponding topic in the manual (marked with the 

 symbol)

before using the relevant function.

Indicates a double-insulated device.

Indicates DC (Direct Current).

Indicates AC (Alternating Current).

Indicates both DC (Direct Current) and AC (Alternating Current).

Indicates a grounding terminal.

Indicates that the instrument may be connected to or disconnected from a
live circuit.

This symbol indicates that the product conforms to regulations set out by
the EC Directive.

WEEE marking:
This symbol indicates that the electrical and electronic appliance is put on
the EU market after August 13, 2005, and producers of the Member States
are required to display it on the appliance under Article 11.2 of Directive
2002/96/EC (WEEE).

Indicates that incorrect operation presents an extreme hazard that
could result in serious injury or death to the user.

Indicates that incorrect operation presents a significant hazard that
could result in serious injury or death to the user.

Indicates that incorrect operation presents a possibility of injury to the
user or damage to the device.

Indicates advisory items related to performance or correct operation of
the instrument.

Usage Notes

To avoid electric shock, do not touch the portion beyond the
protective barrier during use.

• Check that there is no damage to the clamp sensor, instru-

ment case before using. Do not use if there is any damage as
it could lead to electric shock.

• Before using the instrument, make sure that the insulation on

the test leads is undamaged and that no bare conductors are
improperly exposed. Using the instrument in such conditions
could cause an electric shock, so contact your dealer or Hioki
representative for repair.

• During current measurement, to avoid an electric shock acci-

dent, do not connect the test leads to the instrument.

• To avoid electric shock when measuring live lines, wear

appropriate protective gear, such as insulated rubber gloves,
boots and a safety helmet.

Be careful to avoid dropping the instrument or otherwise subjecting them to
mechanical shock, which could damage the mating surfaces of the core and
adversely affect measurement.

• Keep the clamp jaws and core slits free from foreign objects, which could

interfere with clamping action.

Avoid the following locations that could cause an accident or damage to the instrument.

Exposed to direct sun-
light
Exposed to high tem-
perature

In the presence of corro-
sive or explosive gases

Exposed to liquids
Exposed to high hu-
midity or condensation

Exposed to strong elec-
tromagnetic fields
Near electromagnetic 
radiators

Exposed to high levels 
of particulate dust

Near induction heating sys-
tems. (e.g., high-frequency 
induction heating systems 
and IH cooking utensils)

Subject to vibration

Names and Functions of Parts

Operation grip

The clamp sensor at the same side as the pushed grip

opens.

Function selector

OFF, AC current [

A], DC current [

A], AC Voltage/DC Volt-

age [

 V/

 V], Resistance and continuity test [

/

]

(Power is turned on in any position other than OFF.)

Test lead Plug

Connect the test lead plug to the measurement terminal of the
instrument for voltage measurement, resistance measurement or
continuity testing.

 key

 key

• For  [

/

] function: Resistance measurement [

] or Continuity

testing [

] is switchable.

• For [

V/

V] function: AC voltage [

V] or DC voltage

[

 V] is switchable.

• For [

 A] function: Pressing this key together with 

HOLD

 key

initiates zero adjustment.

HOLD

 key

When press 

HOLD

 key, 

 appears in the display and the digi-

tal display value is maintained. Press 

HOLD

 key again to cancel

the data HOLD function.

Functions and Display

The auto power save function cannot be canceled. A minute amount of power con-
tinues to flow while in the power save state. If you will not be using the tester for an
extended period of time, set the function selector to OFF or remove the battery.

LCD panel

Clamp sensor

Current direction indicator

Operation grip

Function selector

HOLD

 key

Measurement 

Plug

key

*Battery cover (rear side)

Test lead plug

Barrier

key

(Display: Measurement value,

Units, Symbols, Decimal point)

terminal

Red test lead (+)

Black test lead (-)

Sleeves

Please do not perform zero adjustment while there is any input to the instrument.
Also note that the zero-adjust function will not function when the display count is
greater than 1000.

Specifications

Zero-adjust Function

Before measuring DC current [

A], you must perform

zero adjustment by simultaneously pressing the 
and 

HOLD

 keys.

Data hold indication

 

Auto power save

Power save state when 10 minutes have elapsed since
the last operation.

LCD panel

4199 maximum display value 

Out of range indication

O.F or -O.F 

Battery low warning 

” is on, the measurement accuracy cannot be guaranteed.

Zero suppression

5 count or less (current measurement only) 

Display update rate

400 ms ±25 ms 

Range switching

Auto range / Manual range 

Crest factor

Current range: 3 max. (1000 A range is 2 max.)
Voltage range: 1.5 max.

Dielectric strength

4290 V rms sine wave ( for 1 min ) between case and cir-
cuit.
7060 V rms sine wave ( for 1 min ) between clamp sensor
and case.
5400 V rms sine wave ( for 1 min ) between clamp sensor
and circuit.

Location for use

Altitude up to 2000 m (6562 feet), indoors
Pollution Degree 2 

Maximum conductor diameter
for measurement 

35 mm (1.38") or less

Operating temperature and
humidity

0 to 40°C (32 to 104°F), 80%RH max. (no condensation) 

Temperature characteristics

In 0 to 40°C range: 0.1 x Measurement accuracy /°C
(In 32°F to 104°F range: 0.56 x Measurement accuracy /°F)

Storage temperature

-10 to 50°C (14 to 122°F), 80%RH max. (no condensation)

Power supply

CR2032 lithium battery x 1 (Rated supply voltage 3 V DC) 

Maximum rated power

15 mVA 

Continuous operating time

Approx. 35 hours (continuous, no load) 

Dimensions 

Approx. 57W x 180H x 16D mm (2.24"W x 7.09"H x 0.63"D)

Mass

Approx. 150 g (5.3 oz.)

Accessories

Instruction Manual, L9208 Test Leads, 9398 Carrying Case

Maximum rated voltage to earth

Current measurement (ACA, DCA):

Measurement Category CAT III 600 V
(anticipated transient overvoltage 6000 V)

Voltage measurement (ACV, DCV):

Measurement Category CATIII 300 V, CATII 600 V
(anticipated transient overvoltage 4000 V)

Applicable standards

Safety : EN 61010, EMC: EN 61326 

Function

Range

Accuracy   ±(%rdg.+dgt.)*

Max. input current

45 to 66 Hz

10 to 45, 66 to 500 Hz

ACA 

( A) 

100.0 A

1000 A 

±(1.5%+5)

±(2.0%+5)

1000 A rms AC

continuous

(See 

Figure

. 1)

DCA

( A)

DC ±(1.5%+5)

1000 A DC 
continuous 

Effect of conductor position: within ±2.0% (in any direction from sensor center)