4.4 Setting Event Detection [REC&EVENT]
68
By enabling Start/Stop Events, measurements are recorded as an event
each time recording is started and stopped.
Manual event recording records the measurement state at the current time
as an event.
Press the
key to display the [REC&EVENT] screen.
Timed Event
Records events at specified
time intervals.
Voltage Swell, Dip and
Interruption
Any momentary deviation
of rms voltage from the nom-
inal voltage is recorded as an
event.
Opening the [REC&EVENT] Screen
Inrush Event
(Inrush, Starting or
Surge Current)
An event is recorded
when the specified value
is exceeded.
Transient Event
(Transient Overvolt-
age)
An event is recorded
when high frequency im-
pulse noise is imposed on
the mains voltage wave-
form.
Start Event/ Stop Event
Manual Events
The time of a manual event is the moment at
which these keys are pressed simultaneously.
Manual event recording is always enabled.
See:
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...