2.4 Common Screen Elements
33
2
7
Current Range
A red field indicates that the crest factor is out
of range.
Select a higher range
9
Normal Line Voltage
The currently selected nominal line voltage is
displayed.
This field is red when the measured voltage
(on channel 1) is far from the selected nomi-
nal line voltage.
11
Internal Operating States (p. 34)
8
PT/CT Ratio
SC
Appears when a PT or CT ratio
has been set.
Nothing appears here when the
PT and CT ratio settings are 1
(1:1).
10
Measured Line Frequency
Red field indicates indicate that the measured
frequency does not match the nominal line
frequency setting.
When no voltage is applied, 0.00 Hz is dis-
played.
The internal operating state changes when you press the
START/STOP
key to start and stop record-
ing data to internal memory.
In order to measure again,
[SET]
must be displayed.
To return to the
[SET]
state, press the
DATA RESET
key, and select whether to save or erase re-
corded data.
Internal Operation Indication
Internal State Description
Data
Recording
Settings
Available
[SET]
Appears after turning power on, until
recording starts.
Before
Recording
All Available
[WAITING]
When a preset start time has been
set, appears while waiting for the re-
cording start time after pressing Start.
Before
Recording
Some Not
Available
[RECORD-
ING]
Appears when recording starts, and
until measurement data has finished
being saved to internal memory.
Recording
Some Not
Available
[ANALYZE]
Appears after recording is finished,
indicating that the data stored in the
internal memory is ready for analysis
Recording
Stops
Some Not
Available
[REVIEW]
Indicates that the data recorded to in-
ternal memory is ready for analysis.
Recorded
(Other data)
Some Not
Available
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...