105
Chapter 6 Maintenance and Service
To clear the error indication, press any key.
6.4 Error Indications and Corrective
Action
Error
Indication
Explanation
Corrective Action
Error at start of re-
cording in Overwrite
Recording Mode
(CONT)
Indicates that memo-
ry already contains re-
corded data.
When recording in Over-
write Recording Mode
(CONT) is desired, delete
all data in memory.
When not desired to de-
lete recorded data in
memory, execute record-
ing in Automatic Stopping
Mode.
Error at start of re-
cording
Indicates that there is
no empty memory
block.
If recording is still desired,
delete all data in memory
(this deletes all data).
Download the necessary
recorded data to a PC be-
fore deleting it.
Data download (page 73)
Error at execution of
memory deletion or
switching from the
monitoring screen
[MONI] to the data
view screen [VIEW]
Indicates that there is
no data in memory.
Memory deletion is not
executed. Switching to
the data view screen
[VIEW] is invalidated.
Other error
Some sort of malfunction
is suspected.
Conduct self-tests (see
If the above does not
clear the error indication,
contact your dealer or
Hioki representative.
Содержание 3144-20
Страница 4: ...Contents iv ...
Страница 14: ...Operating Precautions 10 ...
Страница 28: ...Chapter 1 Overview 24 ...
Страница 46: ...Chapter 2 Measurement Preparations 42 ...
Страница 96: ...Chapter 4 Analysis of Recorded Data 92 ...
Страница 117: ......
Страница 118: ......